Step 1:
- Remove sample system from sterile package.
- Apply a gasket, Three60 unit and clamp to flange interface of your vessel (1 1/2” sanitary connector).
- Complete your “steam in place” or “clean in place” process.
Step 2:
- When ready to remove sample, depress finger tab feature while turning main body clock-wise.
- Immediately release finger tab, as the tab will lock when the open flow position is reached (holding tab as you turn may allow turning past the flow opening).
- Fill product to the volume desired.
Step 3:
- When sample fill is complete, press finger tab and again grip and advance body clock-wise.
- Release tab as you turn as this will assure you lock into the next closed position.
Step 4:
- Initially grip the cutting clamps and press down until the features click.
- When you are sure the lock has engaged, press thumbs together. This action will break the clamps to allow separation.
- Twist the components 360° in opposite directions while holding the components together.
- Pull clamps apart, ripping the tubing. A short extension of tubing will stay attached to the sampling flange, and the bag can be taken to the lab for testing.
|